Viscosity Process Analyzer VISC-4
VISC-4
Funktsioonid
- ASTM compliance
- Capillary type viscometer
- Kinematic viscosity directly measured
- Calculation of dynamic viscosity
- Hagenbach correction not necessary
- Customized solutions
- Network and Fieldbus communication
- Direct and continuous measurement of kinematic viscosity
- Direct comparison with laboratory results without any need for conversion
- Integral measurement of the density
- Calculation and display of the dynamic viscosity
- Temperature control and insulating system without oil bath/pumps
- Minimized maintenance requirements
- Compliance of the temperature stability (0.02 K) as defi ned in standard ASTM D 445
- Necessity of Hagenbach correction is elminated
- Multi-stream capability
- Automatic rinsing and draining facility
- Integrated failure diagnosis and self monitoring
- No atmospheric drain required, backpressure at analyzer outlet permitted
- Single-Phase Power Supply
- Wide range of acceptable sample- and coolant temperature at analyzer inlet
Ex kaitse liik
depending on measuring temperature
Ümbritseva keskkonna temperatuur
DIN EN ISO 3104
IP 71
Kirjeldus
Application:The BARTEC BENKE Viscosity Process Analyzer
(VISC-4) is a continuously kinematic viscosity measuring capillary analyzer suited to measure the viscosity of a product.
The VISC-4 operates online. It serves to monitor/maintain product quality for the in-spec production of mineral oil products.
Three basic variants are available:
- measuring temperature 20 to 60?C (68 to 140?F)
- measuring temperature 41 to 60?C (106 to 140?F)
- measuring temperature 61 to 100?C (142 to 212?F)
Each variant available with following measuring ranges:
- viscosity 0.7 to 30 cSt
- viscosity 10 to 500 cSt
Method:
The value of kinematic viscosity results from the product of the measured passage time and the device constant of the capillary:
v = C * t.
As viscosity is highly dependent on the temperature, the temperature of the liquid during the measurement has to be regulated precisely with minimum variation (0.02 K in accordance with ASTM D 445). For a continuous determination of the viscosity of a liquid during the production process the liquid is controlled in viscometers through a capillary.
From the mass fl ow rate through the capillary and from the pressure drop over the length of the capillary, the current value of the kinematic viscosity of the liquid is ascertained by applying the law of Hagen-Poiseuille.
Tehnilised andmed (väljavõte)
(Täpsema info jaoks klikka siia)Protection class | IP 54, (NEMA 12) |
Dimensions | (W x H x D) approx. 1190 x 1930 x 710 mm |
Weight | approx. 250 kg (without options) |
Pipe fittings | Swagelok? 6 mm/12 mm (other fittings on request) |
Measurement range | L TM : 20 to 60 ?C (68 to 140 ?F) M TM : 41 to 60 ?C (106 to 140 ?F) H TM : 61 to 100 ?C (142 to 212 ?F) t viscosity 0.7 to 30 cSt v viscosity 10 to 500 cSt |
Display | TFT display with touch function (800 x 600 pixels) |
Keyboard | virtual keyboard, controlled via TFT display with touch function |
Operating system | Windows XP? |
Voltage | AC 230 V ? 10 %, 1 phase; 50 Hz (other ratings on request) |
Power | max. approx. 500 W |